CDSCO Licence for Dialysers in India: Class C Registration and ISO 8637
A dialyser is the artificial kidney at the heart of haemodialysis — the cartridge of hollow-fibre membranes that filters a patient’s blood when their kidneys cannot. Because it is a blood-contacting device central to a life-sustaining therapy, CDSCO regulates the dialyser as a Class C device with the haemodialyser standard, sterility and biocompatibility behind the licence. This guide covers the full CDSCO registration pathway for dialysers under the Medical Devices Rules, 2017 — classification, licence route, ISO 8637, the documents, the fees and the timeline. This guide explains CDSCO registration for dialysers in India — risk class, licence forms, standards, documents, fees and timeline.
Is a CDSCO Registration Mandatory for Dialysers?
Yes. A dialyser is a notified Class C medical device and cannot be manufactured, imported or sold in India without a CDSCO licence. Because it is Class C, a domestic manufacturer applies to the Central Licensing Authority in Form MD-7 for a Form MD-9 licence. An importer applies in Form MD-14 for a Form MD-15 import licence. Dialysers, and the associated blood tubing sets and concentrates as their own devices, fall within the framework.
Risk Classification of Dialysers
Dialysers are classified as Class C (moderate-to-high risk). A dialyser is a blood-contacting device essential to a life-sustaining therapy, which places it in Class C (moderate-to-high risk) under the First Schedule. Where a device could fall in two classes the higher class applies, and an accessory is classified in its own right. Always confirm against the current CDSCO classification list.
Which CDSCO Licence You Need for Dialysers
- To manufacture dialysers in India (Class C) — apply in Form MD-7 to the Central Licensing Authority (CDSCO headquarters). The licensing authority grants the licence in Form MD-9. A CDSCO joint audit team inspects the manufacturing site before the licensing authority grants the licence — site inspection is mandatory for Class C and D.
- To import dialysers (any class) — the overseas manufacturer appoints an Indian Authorised Agent who files Form MD-14 with the Central Licensing Authority; CDSCO grants the import licence in Form MD-15.
Applicable Standards for Dialysers
Dialysers are evaluated against the haemodialyser standard plus sterility and biocompatibility:
- ISO 8637-1 — extracorporeal blood circuit: haemodialysers, haemodiafilters, haemofilters and haemoconcentrators.
- ISO 8637-2 — extracorporeal blood circuit for haemodialysers (bloodlines), where relevant.
- ISO 10993 series — haemocompatibility and biological evaluation of the blood-contacting membrane.
- ISO 11135 / 11137 — sterilisation validation (EO, radiation or steam).
- Clearance and ultrafiltration performance data (urea, creatinine, phosphate, KoA).
- ISO 23500 series — where dialysis fluid/water quality is in scope.
Documents Required for CDSCO Registration of Dialysers
The core CDSCO dossier is the same for every medical device: the application on CDSCO’s online medical device portal, Device Master File, Plant Master File, ISO 13485 certificate, Essential Principles checklist, ISO 14971 risk file, labels and IFU, undertakings and the fee receipt. The full checklist, with what each document must contain, is in our CDSCO medical device registration guide. Below is what CDSCO specifically expects to see for dialysers.
Device-Specific Documents for Dialysers
Device-specific documents on top of the core set for dialysers:
- ISO 8637 performance reports — clearance, ultrafiltration coefficient, priming volume and integrity.
- Haemocompatibility (ISO 10993) reports for the membrane and potting materials.
- Sterilisation validation with the sterility assurance level and residual limits.
- Reuse instructions where a reprocessing claim is made, with validation.
- Shelf-life / stability data supporting the expiry.
Label & IFU Requirements Specific to Dialysers
Dialyser labelling is checked for membrane, surface area and sterility:
- Membrane type, surface area and the exact reference number.
- Batch/lot number, date of manufacture and use-by date.
- “STERILE” with the sterilisation-method symbol and the sterilant residual information.
- A single-use symbol (or validated reuse instructions) and priming-volume data in the IFU.
Dialyser, Bloodlines and Concentrate: Separate Devices
Haemodialysis relies on several distinct devices — the dialyser cartridge, the blood tubing set (bloodlines), the dialysis concentrate/fluid and the dialysis machine. They are regulated separately, each with its own class and evidence, even though they are used together. Register the dialyser on its own performance and haemocompatibility evidence, and treat the bloodlines and concentrate as their own applications. Bundling them without distinct evidence is a common cause of queries.
Step-by-Step CDSCO Registration Process for Dialysers
Every device follows the same online filing route, which we walk through step by step in our complete CDSCO registration guide. For dialysers, these are the steps that need the most attention:
- Confirm the classification. Fix the risk class against the CDSCO classification list and the First Schedule — Class C here — which decides the form, authority, fee and audit route.
CDSCO Registration Fees and Timeline for Dialysers
A complete Class C manufacturing file typically clears in about six to nine months. An import (MD-15) file usually takes six to nine months. Every deficiency letter effectively restarts the clock. Unlike the old registration certificates, an MDR 2017 licence does not expire — it stays valid as long as you pay the retention fee (equal to the licence fee) every five years and meet the conditions of licence.
Common Mistakes That Trigger a CDSCO Query
Class C blood-contacting devices draw detailed queries. The recurring reasons are:
- Clearance/ultrafiltration performance data missing or incomplete.
- Haemocompatibility not covering the membrane and potting materials.
- Sterilant-residual data absent where EO sterilisation is used.
- Reuse claims without validated reprocessing instructions.
- Free Sale Certificate scope not covering all surface-area variants on import.
FAQs on CDSCO Registration for Dialysers
What CDSCO class is a dialyser?
A dialyser is Class C, so a domestic manufacturer files Form MD-7 for an MD-9 licence with the Central Licensing Authority, and an importer files Form MD-14 for an MD-15 licence.
Which standard applies to dialysers in India?
ISO 8637-1 for haemodialysers (and ISO 8637-2 for bloodlines), with ISO 10993 haemocompatibility and validated sterilisation.
Does CDSCO require a site inspection?
Yes — for Class C manufacture, a CDSCO joint audit team inspects the site before the licensing authority grants the MD-9 licence.
Are bloodlines and concentrate registered separately?
Yes — the dialyser, blood tubing sets and dialysis concentrate are distinct devices, each with its own class and evidence, even though used together.
How long does dialyser registration take?
Budget roughly six to nine months for a complete Class C manufacturing (MD-9) or import (MD-15) file with performance and haemocompatibility evidence.
Related CDSCO Registration Guides
Need end-to-end help? See our CDSCO registration services, or read these related guides:
- CDSCO Registration for Condoms in India
- CDSCO Registration for Adhesive Bandages in India
- CDSCO Registration for Coronary Stents in India
- CDSCO Registration for Contact Lenses in India
Official references: Central Drugs Standard Control Organisation (CDSCO) and the CDSCO MD Online portal.
Why Choose PCN India Global
- Classification triage — we confirm the correct risk class for your dialysers against the current CDSCO list before we draft anything.
PCN India Global manages the complete CDSCO process for dialysers — from classification and route selection to Device and Plant Master File preparation, online filing, audit support and the licence grant. WhatsApp or call +91 80109 05029, or email admin@pcnindiaglobal.com, to get your registration underway.
Related Compliance Guides
- CDSCO Licence for Coronary Stents
- CDSCO Registration for Syringes
- CDSCO Licence for Surgical Gloves
- CDSCO Licence for Ventilators
Need help? PCN India Global handles end-to-end CDSCO medical device registration for manufacturers and importers.
Related guides
- CDSCO Licence for Digital Thermometers in India: Class B Registration and ISO 80601-2-56
- CDSCO Licence for ECG Machines in India: Class B Registration and IEC 60601-2-25
- CDSCO Licence for Electrosurgical Units in India: Class C Registration and IEC 60601-2-2
- CDSCO Licence for Endoscopes in India: Class C Registration and ISO 8600
- CDSCO Licence for Endotracheal Tubes in India: Class B Registration and ISO 5361
- CDSCO Licence for Enteral Feeding Tubes in India: Class B Registration and ISO 20695
More Medical Devices Guides
- CDSCO Licence for Digital Thermometers in India: Class B Registration and ISO 80601-2-56
- CDSCO Licence for Dental Implants in India: Class C Registration and ISO 14801
- CDSCO Licence for ECG Machines in India: Class B Registration and IEC 60601-2-25
- CDSCO Licence for Dental Chairs in India: Class B Registration and ISO 6875
- CDSCO Licence for Electrosurgical Units in India: Class C Registration and IEC 60601-2-2
- CDSCO Licence for Defibrillators in India: Class C Registration and IEC 60601-2-4