CDSCO Licence for Autoclaves & Sterilizers in India: Class B Registration, ISO 17665 and Process

CDSCO Licence for Autoclaves & Sterilizers in India: Class B Registration and ISO 17665

A steam sterilizer (autoclave) makes reusable instruments safe to use again, so its cycle performance underpins infection control across every hospital and clinic. CDSCO regulates the medical sterilizer as a Class B device with the moist-heat sterilization standards and cycle validation behind the licence. This guide covers the full CDSCO pathway for autoclaves and sterilizers under the Medical Devices Rules, 2017 — classification, licence route, ISO 17665 and EN 285, the documents, the fees and the timeline. (Confirm the exact class for large/industrial sterilizers against the current CDSCO list.)

Is a CDSCO Licence for Autoclaves & Sterilizers Mandatory in India?

Yes. A medical sterilizer is a notified Class B medical device and cannot be manufactured, imported or sold in India without a CDSCO licence. A domestic manufacturer holds a Form MD-5 licence from the State Licensing Authority; an importer holds a Form MD-15 import licence through an Indian Authorised Agent. Tabletop, front-loading and large steam sterilizers fall within the framework.

Risk Classification of Autoclaves & Sterilizers

Every medical device in India is placed in one of four risk classes (A, B, C, D) under the First Schedule of the Medical Devices Rules, 2017. The class fixes the authority, the form, the fee, the audit route and the timeline.

Autoclaves & Sterilizers are classified as Class B (low-to-moderate risk). A sterilizer is a device whose failure leaves instruments non-sterile and creates an infection risk, but which does not itself contact the patient, placing it in Class B (low-to-moderate risk) under the First Schedule. Where a device could fall in two classes the higher class applies, and an accessory is classified in its own right; always confirm against the current CDSCO classification list.

CDSCO Licence for Autoclaves & Sterilizers: Which Licence You Need

  • To manufacture autoclaves & sterilizers in India (Class B) — apply in Form MD-3 to the State Licensing Authority (SLA); the licence is granted in Form MD-5. A CDSCO-empanelled Notified Body audits the site. For Class A the audit is post-grant; for Class B it is carried out before the licence is granted.

Applicable Standards for Autoclaves & Sterilizers

Autoclaves are evaluated against the moist-heat sterilization standards plus equipment and pressure safety:

  • ISO 17665 — sterilization of health-care products: moist heat: requirements for development, validation and routine control.
  • EN 285 — sterilization: steam sterilizers: large sterilizers (or EN 13060 for small steam sterilizers).
  • IEC 61010-2-040 — safety requirements for sterilizers using steam.
  • IEC 60601-1 / IEC 60601-1-2 — where the sterilizer is treated as medical electrical equipment.
  • Pressure-vessel safety conformity (and, in India, relevant statutory pressure-vessel requirements).
  • Cycle-validation and Bowie-Dick / biological-indicator evidence.

Documents Required for Autoclaves & Sterilizers CDSCO Licensing

The core CDSCO dossier is the same for every medical device: the application on CDSCO’s online medical device portal, Device Master File, Plant Master File, ISO 13485 certificate, Essential Principles checklist, ISO 14971 risk file, labels and IFU, undertakings and the fee receipt. The full checklist, with what each document must contain, is in our CDSCO medical device registration guide. Below is what CDSCO specifically expects to see for autoclaves & sterilizers.

Label & IFU Requirements Specific to Autoclaves & Sterilizers

Autoclave labelling and the IFU are checked for cycle, safety and pressure information:

  • Device name, model and chamber volume, and rated cycle parameters.
  • Rated electrical/steam supply and pressure rating with safety symbols.
  • Batch/serial identification and the date of manufacture.
  • Cycle selection, loading and validation/routine-monitoring information in the IFU.

Cycle Validation Is the Product

An autoclave is judged less on its hardware than on its cycles: CDSCO expects moist-heat cycle validation to ISO 17665 and type-test evidence (EN 285 for large, EN 13060 for small sterilizers), plus routine-monitoring aids such as Bowie-Dick tests and biological indicators. Because it is a pressure vessel, statutory pressure-vessel safety also applies. Decide the sterilizer size and cycles, and get the validation and pressure-safety evidence right — a sterilizer without validated cycles is an incomplete file.

Step-by-Step: CDSCO Registration Process

Every device follows the same online filing route, which we walk through step by step in our complete CDSCO registration guide. For autoclaves & sterilizers, these are the steps that need the most attention:

  • Confirm the classification. Fix the risk class against the CDSCO classification list and the First Schedule — Class B here — which decides the form, authority, fee and audit route.
  • Site audit / inspection. A CDSCO-empanelled Notified Body audits the site. For Class A the audit is post-grant; for Class B it is carried out before the licence is granted.

Government Fees and Timeline

Fees are set in the Second Schedule of the Medical Devices Rules and paid online through the CDSCO portal. For manufacture of a Class B device the fee is Rs 5,000 per site plus Rs 500 for each distinct device, paid to the State Licensing Authority (SLA). Import fees under Form MD-15 are charged per site and per device in USD — roughly USD 1,000 per site plus USD 50 per device for lower-risk devices, up to about USD 3,000 per site plus USD 1,500 per device for Class C and D. A test licence (Form MD-13) carries a modest fee. Always confirm the current figure on the CDSCO portal before filing.

A complete Class B manufacturing file is typically cleared in about three to six months; an import (MD-15) file usually takes six to nine months. Every deficiency letter effectively restarts the clock. Unlike the old registration certificates, an MDR 2017 licence does not expire — it stays valid as long as the retention fee (equal to the licence fee) is paid every five years and the conditions of licence are met.

Common Mistakes That Trigger a CDSCO Query

Common query reasons for autoclave applications are:

  • Cycle-validation (ISO 17665) evidence missing or incomplete.
  • Wrong type-test standard for the sterilizer size (EN 285 vs EN 13060).
  • Steam/electrical-safety (IEC 61010-2-040) reports absent.
  • Pressure-vessel conformity not documented.
  • Free Sale Certificate scope not covering the model on import.

Frequently Asked Questions

What CDSCO class is an autoclave/sterilizer?

A medical steam sterilizer is Class B. A domestic manufacturer files Form MD-3 for an MD-5 licence; an importer files Form MD-14 for an MD-15 licence. Confirm the class for large/industrial sterilizers on the current CDSCO list.

Which standards apply to sterilizers in India?

ISO 17665 for moist-heat sterilization validation, EN 285 (large) or EN 13060 (small) type tests, and IEC 61010-2-040 for steam-sterilizer safety.

What is the key evidence for a sterilizer?

Cycle validation to ISO 17665 with routine-monitoring aids (Bowie-Dick, biological indicators) — a sterilizer without validated cycles will not clear.

Do pressure-vessel rules apply?

Yes — an autoclave is a pressure vessel, so statutory pressure-vessel safety requirements apply alongside the CDSCO device licence.

How long does registration take?

A complete Class B manufacturing (MD-5) file is usually cleared in about three to six months; an import (MD-15) file typically takes six to nine months.

Why Choose PCN India Global

  • Classification triage — we confirm the correct risk class for your autoclaves & sterilizers against the current CDSCO list before anything is drafted.

Related Compliance Guides

  • CDSCO Licence for Surgical Microscopes
  • CDSCO Licence for Blood Gas Analysers
  • CDSCO Licence for Ostomy Bags
  • CDSCO Licence for Enteral Feeding Tubes

Need help? PCN India Global handles end-to-end CDSCO medical device registration for manufacturers and importers.