CDSCO Licence for Blood Collection Tubes in India: Class A Registration and ISO 6710
A vacuum blood collection tube (vacutainer) draws and holds a precise volume of blood with the right additive for each test — the humble but essential first step of most laboratory diagnostics. CDSCO regulates blood collection tubes as a Class A device (a specimen receptacle) with the single-use blood-container standard behind the licence. This guide covers the full CDSCO registration pathway for blood collection tubes under the Medical Devices Rules, 2017 — classification, licence route, ISO 6710, the documents, the fees and the timeline. This guide explains CDSCO registration for blood collection tubes in India — risk class, licence forms, standards, documents, fees and timeline.
Is CDSCO Registration for Blood Collection Tubes Mandatory in India?
Yes. A blood collection tube is a notified Class A medical device (a specimen receptacle) and cannot be manufactured, imported or sold in India without a CDSCO licence. A domestic manufacturer holds a Form MD-5 licence from the State Licensing Authority. An importer holds a Form MD-15 import licence through an Indian Authorised Agent. Serum, EDTA, citrate, heparin and fluoride tubes fall within the framework.
Risk Classification of Blood Collection Tubes
Blood Collection Tubes are classified as Class A (low risk). A blood collection tube is a low-risk single-use specimen receptacle, placing it in Class A under the First Schedule — though its draw volume, additive and sterility must still be evidenced. Where a device could fall in two classes the higher class applies, and an accessory is classified in its own right. Always confirm against the current CDSCO classification list.
CDSCO Licence for Blood Collection Tubes: Which Licence You Need
- To manufacture blood collection tubes in India (Class A) — apply in Form MD-3 to the State Licensing Authority (SLA). The licensing authority grants the licence in Form MD-5. A CDSCO-empanelled Notified Body audits the site. For Class A the audit is post-grant. For Class B the Notified Body conducts it before the licensing authority grants the licence.
- To import blood collection tubes (any class) — the overseas manufacturer appoints an Indian Authorised Agent who files Form MD-14 with the Central Licensing Authority; CDSCO grants the import licence in Form MD-15.
Applicable Standards for Blood Collection Tubes
Blood collection tubes are evaluated against the single-use blood-container standard plus additive and sterility evidence:
- ISO 6710 — single-use containers for human venous blood specimen collection.
- Draw-volume accuracy and vacuum retention over shelf-life.
- Additive identity, quantity and uniformity (EDTA, citrate, heparin, clot activator, gel).
- ISO 11135 / 11137 — sterilisation validation of the tube interior.
- ISO 10993 series — biological evaluation as relevant to the specimen path.
- Closure/stopper integrity and leak performance.
Documents Required for CDSCO Registration of Blood Collection Tubes
The core CDSCO dossier is the same for every medical device: the application on CDSCO’s online medical device portal, Device Master File, Plant Master File, ISO 13485 certificate, Essential Principles checklist, ISO 14971 risk file, labels and IFU, undertakings and the fee receipt. The full checklist, with what each document must contain, is in our CDSCO medical device registration guide. Below is what CDSCO specifically expects to see for blood collection tubes.
Device-Specific Documents for Blood Collection Tubes
Device-specific documents on top of the core set for blood collection tubes:
- ISO 6710 test reports — draw volume, vacuum retention, closure integrity and centrifugation.
- Additive specification and uniformity data for each tube type.
- Sterility validation of the tube interior.
- Shelf-life / stability data supporting the draw-volume and additive claims.
- Draw-order / colour-code conformity.
- Country-of-origin approval / Free Sale Certificate for imports.
Label & IFU Requirements Specific to Blood Collection Tubes
Blood-collection-tube labelling is checked for additive, draw volume and expiry:
- Tube type/additive (with the standard colour code), nominal draw volume and size.
- Batch/lot number, date of manufacture and expiry date.
- Sterility status and storage conditions.
- Draw-order guidance and handling information on the pack/insert.
The Additive Is Half the Device
A blood collection tube is defined as much by its additive as by the tube: EDTA for haematology, citrate for coagulation, heparin for chemistry, fluoride for glucose, clot activator/gel for serum. The additive identity, quantity and uniformity — and the draw-volume accuracy that fixes the blood-to-additive ratio — are core evidence, because a wrong ratio invalidates the test. Register each tube type with its additive evidenced and the correct colour code. This keeps the tested product matched to the sold product.
Step-by-Step CDSCO Registration Process for Blood Collection Tubes
Every device follows the same online filing route, which we walk through step by step in our complete CDSCO registration guide. For blood collection tubes, these are the steps that need the most attention:
- Confirm the classification. Fix the risk class against the CDSCO classification list and the First Schedule — Class A here — which decides the form, authority, fee and audit route.
- Site audit / inspection. A CDSCO-empanelled Notified Body audits the site. For Class A the audit is post-grant. For Class B the Notified Body conducts it before the licensing authority grants the licence.
CDSCO Registration Fees and Timeline for Blood Collection Tubes
A complete Class A manufacturing file typically clears in about three to six months. An import (MD-15) file usually takes six to nine months. Every deficiency letter effectively restarts the clock. Unlike the old registration certificates, an MDR 2017 licence does not expire — it stays valid as long as you pay the retention fee (equal to the licence fee) every five years and meet the conditions of licence.
Common Mistakes That Trigger a CDSCO Query
Common query reasons for blood-collection-tube applications are:
- Draw-volume accuracy or vacuum-retention data missing over shelf-life.
- Additive identity/quantity/uniformity not evidenced.
- Closure/stopper integrity or leak data absent.
- Sterility validation of the tube interior missing.
- Free Sale Certificate scope not covering all tube types on import.
FAQs on CDSCO Registration for Blood Collection Tubes
What CDSCO class is a blood collection tube?
A blood collection tube (vacutainer) is Class A. A domestic manufacturer files Form MD-3 for an MD-5 licence. An importer files Form MD-14 for an MD-15 licence.
Which standard applies to blood collection tubes in India?
ISO 6710 for single-use venous blood specimen containers, covering draw volume, vacuum, closure integrity and additive performance.
Why does the additive matter so much?
The additive (EDTA, citrate, heparin, etc.) and the draw-volume accuracy fix the blood-to-additive ratio. A wrong ratio invalidates the test, so both are core evidence.
Does a Class A device need a factory audit?
For Class A manufacture, the Notified Body audit is generally carried out after the MD-5 licence is granted, but the QMS and evidence must be in place at filing.
How long does registration take?
A complete Class A manufacturing (MD-5) file usually clears in about three to five months. An import (MD-15) file typically takes six to nine months.
Why Choose PCN India Global
- Classification triage — we confirm the correct risk class for your blood collection tubes against the current CDSCO list before we draft anything.
PCN India Global manages the complete CDSCO process for blood collection tubes — from classification and route selection to Device and Plant Master File preparation, online filing, audit support and the licence grant. WhatsApp or call +91 80109 05029, or email admin@pcnindiaglobal.com, to get your registration underway.
Related Compliance Guides
- CDSCO Licence for Surgical Lasers
- CDSCO Licence for Cochlear Implants
- CDSCO Licence for Ultrasound Gel
- CDSCO Licence for Menstrual Cups
Related CDSCO Registration Guides
Need end-to-end help? See our CDSCO registration services, or read these related guides:
- CDSCO Registration for COVID-19 Antigen Test Kits in India
- CDSCO Registration for Contact Lenses in India
- CDSCO Registration for Blood Bags in India
Official references: Central Drugs Standard Control Organisation (CDSCO) and the CDSCO MD Online portal.
Need help? PCN India Global handles end-to-end CDSCO medical device registration for manufacturers and importers.
Related guides
- CDSCO Licence for Blood Gas Analysers in India: Class C IVD Registration
- CDSCO Licence for Blood Glucose Test Strips in India: Class C IVD Registration
- CDSCO Licence for Blood Pressure Monitors in India: Class B Registration and IEC 80601-2-30
- CDSCO Licence for Blood Tubing Sets in India: Class C Registration and ISO 8637-2
- CDSCO Licence for Bone Cement in India: Class D Registration and ISO 5833
- CDSCO Licence for Cardiac Catheters in India: Class C Registration and ISO 10555
More Medical Devices Guides
- CDSCO Licence for Blood Gas Analysers in India: Class C IVD Registration
- CDSCO Licence for Cardiac Catheters in India: Class C Registration and ISO 10555
- CDSCO Licence for Blood Pressure Monitors in India: Class B Registration and IEC 80601-2-30
- CDSCO Licence for Condoms in India: Class B Registration and ISO 4074
- CDSCO Licence for Blood Tubing Sets in India: Class C Registration and ISO 8637-2
- CDSCO Licence for Adhesive Bandages in India